A Discussion about Dosage
By James Carroll
There has been a lot of discussion about the best method to quantify dosage.
However, it is my opinion that:
- There is no agreed method of defining beam area
- Dosage expressed as J/cm² is inadequate
No agreed method
|
So beam area is hard to define and there is no agreement in our industry for defining it.
[I propose 1/e² - will explain this soon].
Dosage expressed as J/cm² is inadequate
"Dosage" is usually calculated as Power / Beam Area x Time = J/cm². However, to consider that dosage should equal J/cm² is, in my opinion, inadequate.
Let me explain:
Assume there is a well-defined beam area and thus a quantifiable dosage.
- A 500mW laser with a beam area of 0.25Cm² used for 20 seconds delivers 40 J/cm²
- A 200mW laser with a beam area of 0.1Cm² used for 20 seconds delivers 40 J/cm²
- A 30mW laser with a beam area of 0.015Cm² used for 20 seconds also 40 J/cm²
Each of these probes apparently apply the same "dosage". However, the total energy delivered is clearly different [10 Joules, 4 Joules and 0.6 Joules respectively].
While dosage appears consistent using J/cm², I suggest that the clinical results would be quite diverse. So I say that J/cm² is an inadequate method of expressing dosage.